Skip to content

Public FDA record

Z-1282-2015: Siemens LANTIS Oncology Information System Servers; allows the radiation therapist to deliver treatment

Recall record Z-1282-2015 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1282-2015
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Mar 16, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential safety risk when using LANTIS server software with operating systems with which it has not been validated or released which can lead to an incorrect treatment to the patient.

Source facts

Action
Siemens sent an Customer Advisory Notice, dated January 21, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were informed the risk associated with use of more recent operating systems with LANTIS Server Software. Customers were warned against the use of more recent operating systems to ensure the LANTIS system remains in the validated configuration. In the interest of safety, we ask you to perform the above preventive measures and inform all affected personnel immediately. Please include this Field Safety Notice in your System Owner Manual chapter "Safety Advisory Letters" where it should remain. The relevant National Competent Authority will be informed of this Field Safety Notice. Please forward this safety information to other organizations that could be affected by this measure. If the device has been sold and therefore no longer in your possession, please forward this customer notice to the new owner. We would also request that you inform us of the identity of the device's new owner where possible. Please observe this notice until the update has been fully completed. We apologize for any inconvenience and thank you for your cooperation. For further questions please call (610) 219-6300.
Initiated
2015-01-21
Root cause
Use error
Terminated
2015-07-30
Distribution
Nationwide Distribution
Recall number
Z-1282-2015
Recall status
Terminated
CDRH record ID
133396
FDA classified
2015-03-16
Recall event number
70384

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗