Open original FDA record ↗Recall record
- FDA identifier
- Z-1282-2019
- Company in FDA record
- Medtronic Inc.
- FDA record creation date
- May 3, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.
Source facts
- Action
- A Field Corrective Action was initiated outside of the U.S. by the recalling firm's global regions on 7/4/2018.
- Initiated
- 2018-07-04
- Root cause
- Software design
- Terminated
- 2022-06-06
- Distribution
- Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Turkey, and South Africa.
- Recall number
- Z-1282-2019
- Recall status
- Terminated
- CDRH record ID
- 169303
- FDA record created
- 2019-05-03
- Recall event number
- 81656