Open original FDA record ↗Recall record
- FDA identifier
- Z-1282-2021
- Company in FDA record
- Hitachi Medical Systems America Inc
- FDA record creation date
- Mar 19, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Source facts
- Action
- On 3/1/2021, Hitachi issued an Urgent Medical Device Correction notice to customers for the Arietta 850 system operating with software version 4.0.2 that utilizes SWE function with the C252 probe. The R&D department of Hitachi Ltd. (Manufacturer) has identified an error in the focus point and transducer aperture settings in the SWE function at the manufacturer facility. This error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
- Initiated
- 2021-03-01
- Root cause
- Software design
- Terminated
- 2021-11-04
- Distribution
- US Nationwide distribution in the state of TX.
- Recall number
- Z-1282-2021
- Recall status
- Terminated
- CDRH record ID
- 186165
- FDA record created
- 2021-03-19
- Recall event number
- 87493