Skip to content

Public FDA record

Z-1283-2012: Philips Easy Diagnost Eleva DRF, Software release 4.0 Fluoroscopic image-intensified Stationary X-Ray Sy

Recall record Z-1283-2012 for Philips Medical Systems Gmbh, DMC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1283-2012
Company in FDA record
Philips Medical Systems Gmbh, DMC
FDA classification date
Mar 23, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Underexposed image due to lag in switching between auxiliaries

Source facts

Action
Philips Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated February 27, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The firm provided a work around for customers to use until a software update can be installed. Customers should contact their local Philips representative for questions regarding this notice.
Initiated
2012-03-09
Root cause
Software design
Terminated
2013-05-23
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Australia, Germany, Indonesia, Malaysia, Saudi-Arabia, Singapore, and Thailand.
Recall number
Z-1283-2012
Recall status
Terminated
CDRH record ID
107921
FDA classified
2012-03-23
Recall event number
61365

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗