Skip to content

Public FDA record

Z-1283-2022: CardioTek EP-TRACER Software V2.x.

Recall record Z-1283-2022 for CardioTek BV. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1283-2022
Company in FDA record
CardioTek BV
FDA record creation date
Jun 23, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device did not pass electrical safety testing for adequate insulation.

Source facts

Action
The firm issued a recall notification to consignees on 04/25/2022 via email. On 05/24/2022 the firm sent an updated field safety notice. The consignee can continue to use the device. The firm will contact the customer to schedule an appointment for the rework. The consignee is to pass on the letter to those that need to be made aware of the notice.
Initiated
2022-04-25
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, NY, NC and the countries of South Korea, Mexico, Kazakhstan, Netherlands, Spain, Germany, Finland, Poland, Israel, Turkey, Chile, Ecuador, Austria, Croatia, Sweden, United Kingdom, Colombia, Italy, Portugal, Canada, Saudi Arabia, Belgium, Slovenia, Brazil, Hungary, Malaysia, Uzbekistan.
Recall number
Z-1283-2022
Recall status
Completed
CDRH record ID
193584
FDA record created
2022-06-23
Recall event number
90154

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗