Z-1284-2022: QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CA

Recall record Z-1284-2022 for Microgenics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1284-2022
Company in FDA record
Microgenics Corporation
FDA record creation date
Jun 24, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.

Source facts

Action
On 01/20/2022, the firm contact via email its international customers regarding that the QMS Tacrolimus Calibrator was recovering between 82 to 94% of its target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay. On 03/10/2022, the firm notified its U.S. customer of the issue. Customers were instructed to: 1)Determine if they are using or have inventory of QMS Tacrolimus Calibrator catalog number 10015573, lot 74175462 or lot 74267412. 2)Discontinue use of affected in-date lots, and transition to new lot(s) where corrections have been made. Lot 74395275 and Lot 74459127 have had the testing correction applied and are currently available in stock to replace any of their remaining inventory of lots 74175462 or 74267412. 3)Dispose of the effected lots in accordance with local, regional, national regulations. 4)Contact your health care professional, as appropriate, for further action. " This notice needs to be passed on to all who need to be aware within your organization or to any organization where the potentially affected calibrator lots have been transferred. " Tacrolimus results reported with these affected calibrator lots do not need to be reanalyzed. " Retain a copy of this letter for your laboratory records. " Call Technical Service if assistance is needed for lot replacement (free of charge). " Please complete the Acknowledgment & Receipt Form attached to this Notification within 5 days of the date of this letter to Thermo fisher Scientific Technical Service. For questions, contact: -FAX: 888-527-8001 (USA only) or +1-510-979-5420 (outside USA) -PHONE: (+1)-800-232-3342 then press option 2, then option 2, or contact your local country representative. -E-mail: TechService.mgc@thermofisher.com
Initiated
2022-01-20
Root cause
Process control
Distribution
U.S.: AL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC, TN, TX, and VA O.U.S.: Bangladesh, Brazil, Canada, Germany, India, Malaysia, Mexico, South Korea, and Thailand
Recall number
Z-1284-2022
Recall status
Open, Classified
CDRH record ID
193848
FDA record created
2022-06-24
Recall event number
90225

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗