Open original FDA record ↗Recall record
- FDA identifier
- Z-1285-2023
- Company in FDA record
- Preat Corp
- FDA classification date
- Mar 24, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Engaging titanium base contains a dimensional condition that allows vertical movement of the titanium base after the prosthetic screw is torqued to the recommended value, which if undetected and installed on the implant, may lead to abutment/screw fracture, which could result in the abutment/screw/crown being ingested by the patient.
Source facts
- Action
- On 2/21/23, recall notices were mailed to customers who were asked to discard or return affected devices to the recalling firm. In addition, in the event affected product was further distributed, customers were asked to inform their customers, or to provide a customer list so the recalling firm could notify these additional customers. The customer service number was provided: 800-232-7732
- Initiated
- 2023-02-21
- Root cause
- Process design
- Distribution
- US Nationwide distribution in the states of TX, OR, UT, NY.
- Recall number
- Z-1285-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199094
- FDA classified
- 2023-03-24
- Recall event number
- 91876