Open original FDA record ↗Recall record
- FDA identifier
- Z-1288-2015
- Company in FDA record
- Ossur H / F
- FDA classification date
- Mar 17, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module can become loose from the foot blade.
Source facts
- Action
- The firm's recall letter states the following instructions: Immediately examine inventory and quarantine product subject to the recall. In addition the firm asks if further distributed to identify the customers and notify them of the recall. The firm states that notification may be enhanced by providing a copy of the letter. The firm states that the recall notice needs to be passed along to all those who need to be aware. The firm also states that if customers are wearing a faulty product that they recommend them scheduling a visit to have the product replaced with a new LP Rotate that is currently available. The firm states that they can also arrange pick-up of defected products. For any questions the firm provides a customer service contact: Jamie Dutter, Senior Customer Service Representative.
- Initiated
- 2015-03-04
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-10-28
- Distribution
- Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,
- Recall number
- Z-1288-2015
- Recall status
- Terminated
- CDRH record ID
- 134240
- FDA classified
- 2015-03-17
- Recall event number
- 70648