Skip to content

Public FDA record

Z-1288-2021: Interventional Fluoroscopic X-Ray System - Product Usage: developed for single and biplane diagnostic ima

Recall record Z-1288-2021 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1288-2021
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Mar 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens has become aware of a potential issue in which the imaging system UPS sporadically forces a shutdown of the imaging system PC without a true power supply problem.

Source facts

Action
On 2/26/2021, Siemens issued an Urgent Medical Device Correction notice to in inform customers of a possible hazard to patients, operators, or other persons and equipment due to a hardware error which affects ARTIS icono /pheno systems in combination with imaging system UPS (Uninterrupted Power Supply) option. This potential hardware issue has no influence on the previous treatment/diagnosis of patients. However, in case of error, planned procedures may have to be terminated and performed on an alternative x-ray system. Based on our investigation, you can continue to safely use your system.
Initiated
2021-02-26
Root cause
Component design/selection
Terminated
2021-08-11
Distribution
US Nationwide distribution in the states of MI, KY, TX.
Recall number
Z-1288-2021
Recall status
Terminated
CDRH record ID
186134
FDA record created
2021-03-22
Recall event number
87479

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗