Open original FDA record ↗Recall record
- FDA identifier
- Z-1288-2021
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Mar 22, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens has become aware of a potential issue in which the imaging system UPS sporadically forces a shutdown of the imaging system PC without a true power supply problem.
Source facts
- Action
- On 2/26/2021, Siemens issued an Urgent Medical Device Correction notice to in inform customers of a possible hazard to patients, operators, or other persons and equipment due to a hardware error which affects ARTIS icono /pheno systems in combination with imaging system UPS (Uninterrupted Power Supply) option. This potential hardware issue has no influence on the previous treatment/diagnosis of patients. However, in case of error, planned procedures may have to be terminated and performed on an alternative x-ray system. Based on our investigation, you can continue to safely use your system.
- Initiated
- 2021-02-26
- Root cause
- Component design/selection
- Terminated
- 2021-08-11
- Distribution
- US Nationwide distribution in the states of MI, KY, TX.
- Recall number
- Z-1288-2021
- Recall status
- Terminated
- CDRH record ID
- 186134
- FDA record created
- 2021-03-22
- Recall event number
- 87479