Open original FDA record ↗Recall record
- FDA identifier
- Z-1289-2014
- Company in FDA record
- Teleflex Medical
- FDA classification date
- Mar 25, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Complaints were received of leakage of insufflation gas through the device, which can cause difficulty or delay in the completion of laparoscopic surgical procedures.
Source facts
- Action
- Teleflex sent an Urgent Medical Device Recall Notification letter dated March 11, 2014 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to return all affected product to Teleflex Medical per the instructions provided in the letter. Customers were asked to complete the enclosed Recall Acknowledgment Form and fax it to 1-866-804-9881, Attn: Customer Service. For questions contact your local sales representative or Customer Service at 1-866-246-6990.
- Initiated
- 2014-03-11
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-08-15
- Distribution
- Worldwide Distribution - USA Nationwide including Puerto Rica and the countries of: Australia, Brazil, China, Canada, Malaysia, Korea, Ireland Singapore and Vietnam.
- Recall number
- Z-1289-2014
- Recall status
- Terminated
- CDRH record ID
- 126122
- FDA classified
- 2014-03-25
- Recall event number
- 67710