Z-1289-2018: DynaCad software consist of an MR Analysis Server software and the viewer workstation software. The MR A

Recall record Z-1289-2018 for Invivo Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1289-2018
Company in FDA record
Invivo Corporation
FDA classification date
Apr 2, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered correctly on the DynaCad Client. The defect causes the pharmacokinetic (PK) color maps to display incorrectly when viewed from remote DynaCAD client computers and could result in visually underestimating calculated Ktrans, Kep, and iAUGC values. The defect also impacts DynaCAD s on-the-fly calculation of Apparent Diffusion Coefficient (ADC) maps. If the ADC values are computed on-the-fly by DynaCAD, the ADC values and colors will also display incorrectly if viewed on a remote DynaCAD client.

Source facts

Action
Invivo sent an Urgent Medical Device Correction - Field Safety Notice letter dated January 24, 2018 to affected customers. The letter identified the affected product, problem and actions to be taken. This document contains important information for the continued safe and proper use of your equipment. Customers were instructed to review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Customers were asked to retain a copy with the equipment Instruction for Use. For questions contact Invivo Customer Support: 1-877-468-4861 Opt. 3
Initiated
2018-01-24
Root cause
Other
Terminated
2020-05-04
Distribution
Device is software. Customer notification letters recommended that users discontinue use of the Ktrans, Kep, and iAUGC colormaps when assessing studies from a remotely connected DynaCAD client computer. It is further advised that users refrain from using the colormaps and values derived from ADC maps calculated by DynaCAD. ADC maps originating natively from the MRI system should be used as an alternative as these values are unaffected. Firm will provide a v4.0 software update for the affected software versions (v3.4, v3.5) to correct the defect at no charge to the user.
Recall number
Z-1289-2018
Recall status
Terminated
CDRH record ID
162733
FDA classified
2018-04-02
Recall event number
79590

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗