Open original FDA record ↗Recall record
- FDA identifier
- Z-1289-2025
- Company in FDA record
- Smiths Medical ASD, Inc.
- FDA classification date
- Mar 12, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Affected devices have a smaller diameter than expected and may potentially result in inadequate ventilation to the patient. Patient may experience hypoxia, underdose, or cardiopulmonary collapse.
Source facts
- Action
- On February 13, 2025, the firm began to notify affected consignees via Urgent Medical Device Correction letters. Customers were informed that affected sizes of endotracheal tube products may have a smaller diameter than expected. Customers were instructed to check all inventory locations for affected products and discontinue use. Affected product should be destroyed following your institution's process for destruction. Product should be quarantined until disposal. All potential users of the device should be informed. If you have distributed affected product, please immediately forward the recall notice to your customers.
- Initiated
- 2025-02-13
- Root cause
- Process change control
- Distribution
- Worldwide - US Nationwide distribution.
- Recall number
- Z-1289-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212450
- FDA classified
- 2025-03-12
- Recall event number
- 96263