Open original FDA record ↗Recall record
- FDA identifier
- Z-1290-2015
- Company in FDA record
- Cardiovascular Systems, Inc.
- FDA classification date
- Mar 20, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The OAD saline sheath outer layer may flake off from the catheter due to friction between the outer material layer and corresponding materials it may come in contact with during clinical use. If the material is large enough, it has the potential to cause a flow limiting embolism.
Source facts
- Action
- Consignees were sent on 3/4/2015 a CSI "Urgent Medical Device Recall" letter dated March 04, 2015. The letter described the problem and the affected product. Consignees were advised to remove and return the affected product to CSI. The letter also requested consignees to complete and return the Customer Acknowledgement Form by Fax or with the RGA (Returned Goods Authorization). For further information they can contact Customer Service, Cardiovascular Systems, Inc., 1225 Old Hwy 8, New Brighton, MN 55112, 877-274-0901 -Tel, 612-677-3355- Fax.
- Initiated
- 2015-03-04
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-04-26
- Distribution
- Nationwide Distribution-including the states of AZ, FL, GA, LA, MA, MS, MO, NJ, NY, OH, and RI.
- Recall number
- Z-1290-2015
- Recall status
- Terminated
- CDRH record ID
- 134565
- FDA classified
- 2015-03-20
- Recall event number
- 70727