The size indicated on the labeling on the outer packaging, on the inner packaging, and on the patient stickers inside the packaging is incorrect: the Trinity Biolox Delta Ceramix Head 32XL is labeled as 36XL and vice-versa.
Source facts
Action
The firm contacted their consignee by email on 01/25/2021. The distributor was instructed to identify the affected product in its inventory and return it to Corin USA in order to contain the part. A recall letter will be sent and acknowledgement of receipt will be requested.
Initiated
2021-01-25
Root cause
Labeling mix-ups
Terminated
2022-11-23
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Colombia, Poland, Germany, Australia.
Recall number
Z-1290-2021
Recall status
Terminated
CDRH record ID
185737
FDA record created
2021-03-22
Recall event number
87293