Open original FDA record ↗Recall record
- FDA identifier
- Z-1290-2024
- Company in FDA record
- Biosense Webster, Inc.
- FDA classification date
- Mar 13, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a manufacturing issue, device under process validation phase were inadvertently mixed into the main manufacturing process.
Source facts
- Action
- On 02/16/2024, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via FedEx to customers informing them that Biosense Webster identified through a complaint investigation that a mix-up occurred during the manufacturing of two lots. The devices could potentially be shorter in length (24 1/8 inches (61.28 cm) instead of, 28 inches (71 cm). Customer are instructed/informed: If a CARTO VIZIGO" Bi-Directional Guiding Sheath, Medium Curve (Catalog Number: D-1385-02) from Lot Numbers 60000282 and 60000287 is being used in their facility, they should continue to follow the IFU. The Instructions for Use, on Page 7, instructs users to inspect all components before use, and prior to inserting the device into the patient, pre-assemble the sheath, dilator, and stylet on the table. While setting up for a procedure, if the dilator is protruding out the tip of the sheath more than 4 inches (10 cm), do not use the sheath. Destroy the device following their standard process. Please complete the Attestation of destruction form (Attachment 2) to receive a replacement and email it to: Email: OneMD-Field-Actions@its.jnj.com Subject: Attention: BWI Field Action Coordinator- Attestation product destruction. For questions or further assistance, contact Biosense Webster at OneMD-Field-Actions@its.jnj.com
- Initiated
- 2024-02-16
- Root cause
- Under Investigation by firm
- Distribution
- U.S. Nationwide distribution including in the states of FL, IL, AL, NY, PA, SC, AZ, MI KY, NJ, WI, IN, LA, TX, MA, OH, AR, MT, MO, VA, NC, TN, CA, MN, WV, CT, CO, IA, MD, ID, NM, NH, UT, GA, WA, NV, OK, KS, AK, NE, HI.
- Recall number
- Z-1290-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206118
- FDA classified
- 2024-03-13
- Recall event number
- 94014