Z-1290-2024: CARTO VIZIGO Bi-Directional Guiding Sheath, REF D128502 The CARTO VIZIGOTM Bi-Directional Guiding Sheat

Recall record Z-1290-2024 for Biosense Webster, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1290-2024
Company in FDA record
Biosense Webster, Inc.
FDA classification date
Mar 13, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue, device under process validation phase were inadvertently mixed into the main manufacturing process.

Source facts

Action
On 02/16/2024, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via FedEx to customers informing them that Biosense Webster identified through a complaint investigation that a mix-up occurred during the manufacturing of two lots. The devices could potentially be shorter in length (24 1/8 inches (61.28 cm) instead of, 28 inches (71 cm). Customer are instructed/informed: If a CARTO VIZIGO" Bi-Directional Guiding Sheath, Medium Curve (Catalog Number: D-1385-02) from Lot Numbers 60000282 and 60000287 is being used in their facility, they should continue to follow the IFU. The Instructions for Use, on Page 7, instructs users to inspect all components before use, and prior to inserting the device into the patient, pre-assemble the sheath, dilator, and stylet on the table. While setting up for a procedure, if the dilator is protruding out the tip of the sheath more than 4 inches (10 cm), do not use the sheath. Destroy the device following their standard process. Please complete the Attestation of destruction form (Attachment 2) to receive a replacement and email it to: Email: OneMD-Field-Actions@its.jnj.com Subject: Attention: BWI Field Action Coordinator- Attestation product destruction. For questions or further assistance, contact Biosense Webster at OneMD-Field-Actions@its.jnj.com
Initiated
2024-02-16
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution including in the states of FL, IL, AL, NY, PA, SC, AZ, MI KY, NJ, WI, IN, LA, TX, MA, OH, AR, MT, MO, VA, NC, TN, CA, MN, WV, CT, CO, IA, MD, ID, NM, NH, UT, GA, WA, NV, OK, KS, AK, NE, HI.
Recall number
Z-1290-2024
Recall status
Open, Classified
CDRH record ID
206118
FDA classified
2024-03-13
Recall event number
94014

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗