Skip to content

Public FDA record

Z-1290-2025: (1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF

Recall record Z-1290-2025 for LUMENIS, LTD.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1290-2025
Company in FDA record
LUMENIS, LTD.
FDA classification date
Mar 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.

Source facts

Action
The recall was initiated via letter dated and issued on 1/23/2025 via FEDEX in the U.S. and via mail, courier, or hand delivery OUS. The letter explained the issue and the scenarios as to when the laser system would not power on. The consignee is instructed to immediately discontinue use and segregate the device, post this information in a visible location near the affected product to ensure the information is readily accessible to all handlers and users of the device, share the notice with any health care professional within your organization and with any organization where the affected product may have been transferred, and if the device has been further distributed, notify the customer to ensure this product removal is carried out to the end-user level. A Reply Verification Tracking Form was enclosed for completion and return. Once the firm receives the completed form, the consignee will be contacted within 2 business days and provided an RGA number for product return.
Initiated
2025-01-23
Root cause
Under Investigation by firm
Distribution
Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.
Recall number
Z-1290-2025
Recall status
Open, Classified
CDRH record ID
212293
FDA classified
2025-03-05
Recall event number
96197

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗