Skip to content

Public FDA record

Z-1292-2021: Duodenoscope and accessories, flexible/rigid - Product Usage: used with an Olympus video system center, l

Recall record Z-1292-2021 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1292-2021
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Mar 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Olympus Medical Systems Corporation (OMSC) has received complaints of mucosal injury associated with the forceps elevator region of the TJF-Q190V duodenoscope, including the distal cover (model MAJ- 2315).

Source facts

Action
On 2/5/2021, Olympus issued Urgent Device Correction notices to customers via letter informing them Olympus has identified an increase in the number of complaints from customers reporting that mucosal tissue was found within the single-use distal cover, MAJ-2315, after withdrawing the TJF-Q190V duodenoscope from the patient. In the reported complaints, the user facilities have reported mucosal injuries in the esophagus, stomach, and duodenum. Olympus is thoroughly investigating the root cause or causes of these reported events. This notice included information that may reduce the potential occurrence of similar incidents.
Initiated
2021-01-28
Root cause
Under Investigation by firm
Terminated
2024-03-13
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Japan, Europe, Australia.
Recall number
Z-1292-2021
Recall status
Terminated
CDRH record ID
185744
FDA record created
2021-03-22
Recall event number
87296

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗