Z-1292-2022: ProudP Everyday Uroflow Tracker
Software Version 1.x.x
Software Version 1.2.0 or earlier
Recall record Z-1292-2022 for Dain Technology, Inc. (Subsidiary of Soundable Health, Inc.). View source facts, labeled dates, and the original FDA record.
Dain Technology, Inc. (Subsidiary of Soundable Health, Inc.)
FDA record creation date
Jun 27, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may perform automatic processing of the urination sound signal, resulting in lower urination volume and velocity values than expected.
Source facts
Action
On 04/21/2022, Soundable Health/ReedSmith LLP (Dain Technology, Inc.) emailed a Customer Notification informing customers that the firm has implemented a mobile application correction to the proudP application due to an issue regarding use of
proudP mobile application with the iPhone Listen Live feature or hearing aids which could result in interference with the data collected by the proudP application and provide inaccurate readings.
Customer are being informed that an update /correction to the application through the Apple App Store was released on April 21, 2022. If you downloaded your proudP before April 21, 2022, and have not updated the application since then, you will not be able to use proudP after April 21, 2022. Do not use any external devices or features (e.g., hearing aids, Listen Live feature) that activates your iPhone s audioprocessing function when collecting data with proudP. If for any reason you are not able to update the product, please email us at gayeong.kim@soundable.health. Do not use the product if you cannot update the product.
For any questions - contact the Regulatory and Quality Assurance Manager at gayeong.kim@soundable.health
Initiated
2022-04-21
Root cause
Software design
Terminated
2023-09-29
Distribution
Worldwide distribution - US Nationwide. The software app is distributed via the Apple App Store.
Recall number
Z-1292-2022
Recall status
Terminated
CDRH record ID
193649
FDA record created
2022-06-27
Recall event number
90165
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.