Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectivity issues.
Source facts
Action
The firm sent a single page of instructions to customers on August 9, 2024 via email. Customers were informed that tablets should remain plugged into a wall outlet when the tablet is not in use. The tablet battery should not be allowed to fully deplete at any time during patient monitoring or when not in use.
Initiated
2024-08-09
Root cause
Labeling design
Distribution
US Nationwide distribution in the states of Florida, Louisiana, Maryland, Montana, New York, Ohio, and Texas.
Recall number
Z-1292-2025
Recall status
Open, Classified
CDRH record ID
212419
FDA classified
2025-03-06
Recall event number
96252
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.