Open original FDA record ↗Recall record
- FDA identifier
- Z-1293-2015
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Mar 18, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Negative drift on QC during in-use lifetime of CREA membranes can be observed.
Source facts
- Action
- Radiometer sent a Product Notification letter dated November 10, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to perform the following steps immediately: " Perform one quality control measurement using S7835 AutoCheck 6+, level 1 on already installed Crea membrane units to verify the Crea performance. If the QC result is out of range, new Crea A and Crea B membranes must be installed. Until further notice the following must be carried out every 24 hours: " Perform one quality control measurement using S7835 AutoCheck 6+, level 1 to verify the Crea performance. If the QC result is out of range, new Crea A and Crea B membranes must be installed. Please do not report Crea results unless the membrane unit has been verified as described above. If you already run S7835 AutoCheck 6+ level 1, every 24 hours, as part of your regular QC schedule, you do not need to do anything additional. Please review your QC results to ensure the results are within the QC range. Please complete and return the attached fax form, with your signature as soon as possible or within 7 days. Customers with questions were instructed to contact Radiometer America Technical Support at 1-800-736-0600, option 4. For questions regarding this recall call 440-871-8900, ext 209.
- Initiated
- 2014-10-29
- Root cause
- Under Investigation by firm
- Terminated
- 2015-07-02
- Distribution
- Worldwide Distribution - USA including NM, NC, TN, OH, NE, CA, MO, GA, MA, TX, WV, PA, NY, VA, MN, IN, ME, MD, IL, MI, KS, WI, FL, WA, CT, OK, CO, and Internationally to Canada, Denmark, Spain, Italy, India, Sweden, Germany, Australia, Hungary, Great Britain, Norway, Uruguay, France, Poland, Switerland, Russian Federation, Netherlands, Czech Republic, Mexico, Brazil, Czechoslovakia, Austria, Singapore, Syria, Phillipines, Egypt, Bolivia, Japan, New Zealand, Turkey, Saudi Arabia, Algeria, Qatar, and Finland.
- Recall number
- Z-1293-2015
- Recall status
- Terminated
- CDRH record ID
- 131577
- FDA classified
- 2015-03-18
- Recall event number
- 69813