Skip to content

Public FDA record

Z-1293-2021: Mizuho CE IPX4, mode MOT-5602BW, 100-240, 50-60Hz, 400VA, REF 18-140-20 containing hand control Unit - Pr

Recall record Z-1293-2021 for Mizuho OSI. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1293-2021
Company in FDA record
Mizuho OSI
FDA record creation date
Mar 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that hand control units may cause the tabletop to slide in the opposite direction from that selected by the user. This could result in a delay in procedure.

Source facts

Action
On 2/3/2021 Mizuho sent a "Recall Notice for Control Unit" notification letter to affected consignees via Email. In addition to informing consignees about the recall the notification asked consignees to take the following actions: 1. Identify if you have the affected product, 2. Mizuho will replace control units. As soon as we are ready for your institutions, Mizuho will supply the replacements. Please use the auxiliary switch on the operating table until the replacements arrive.
Initiated
2021-02-08
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide distribution in the state of CA, and the countries of Costa Rica, Chile, Bolivia, Peru.
Recall number
Z-1293-2021
Recall status
Open, Classified
CDRH record ID
185804
FDA record created
2021-03-23
Recall event number
87318

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗