Open original FDA record ↗Recall record
- FDA identifier
- Z-1293-2021
- Company in FDA record
- Mizuho OSI
- FDA record creation date
- Mar 23, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential that hand control units may cause the tabletop to slide in the opposite direction from that selected by the user. This could result in a delay in procedure.
Source facts
- Action
- On 2/3/2021 Mizuho sent a "Recall Notice for Control Unit" notification letter to affected consignees via Email. In addition to informing consignees about the recall the notification asked consignees to take the following actions: 1. Identify if you have the affected product, 2. Mizuho will replace control units. As soon as we are ready for your institutions, Mizuho will supply the replacements. Please use the auxiliary switch on the operating table until the replacements arrive.
- Initiated
- 2021-02-08
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide distribution - US Nationwide distribution in the state of CA, and the countries of Costa Rica, Chile, Bolivia, Peru.
- Recall number
- Z-1293-2021
- Recall status
- Open, Classified
- CDRH record ID
- 185804
- FDA record created
- 2021-03-23
- Recall event number
- 87318