Z-1293-2023: da Vinci Xi Surgical System Model IS4000 da Vinci X Surgical System Model IS4200

Recall record Z-1293-2023 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1293-2023
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Mar 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to Universal Surgical Manipulator (USM) instrument carriage being loose as a result of the linear rail component not being manufactured within specifications.

Source facts

Action
On 02/10-13/2023, Intuitive Surgical sent "Urgent Medical Device Correction" letters via email to inform two Groups of Customer (Group 1 and Group 2) that, a population of da Vinci Z and Xi Systems instrument arms (USMs-Universal Surgical Manipulators) have instrument Carriages that may separate from the Insertion Axis Linear Rail during shipping, handling, or repositioning of the arms when the system is powered down. The Instrument Carriage would remain attached to the drive motor which prevents free-fall motion or complete detachment from the USM. However, the Instrument Carriage may feel like it is not tightly connected to the rail. This condition may also lead to increased friction along the insertion axis. Instructions for Customers in Group 1 (reference Appendix A of their Letter): Stop/Cease use of the affected systems until Intuitive Representatives can replace the affected USMs on their systems. Instructions for Customers in Group 2 (reference Appendix B of their Letter): Can continue to use affected systems, but prior to each use, need to perform inspection of all the instrument carriages using the "Instrument Carriage Inspection Instructions" (refer letter for details of inspection). For questions or additional guidance: -US, Puerto Rico and Dominican Republic: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com. -Europe, Middle East, Asia, South America and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or EUCS@intusurg.com -South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) or support.korea@intusurg.com -Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) or csjapan@intusurg.com -Taiwan: 886-2-2-27008181 (9 AM to 6:00 PM CST) or CS.taiwan@intusurg.com -India: 1800 103 6952 (9:00 AM to 6:00 PM IST) or Email address: support.india@intusurg.com
Initiated
2023-02-10
Root cause
Equipment maintenance
Distribution
U.S.: Cayman Islands, Dominican Republic, Puerto Rico, United States O.U.S.: Austria, Belgium, Denmark, Finland, France, Germany, Ireland, Luxembourg, Martinique, Monaco, Netherlands, Norway, Reunion, Sweden, Switzerland, United Kingdom, India, Japan, South Korea, Taiwan, Brazil, Argentina, Australia, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Ecuador, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, Indonesia, Israel, Italy, Kuwait, Lebanon, Malaysia, Malta, Mexico, New Zealand, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Thailand, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam
Recall number
Z-1293-2023
Recall status
Open, Classified
CDRH record ID
198066
FDA classified
2023-03-28
Recall event number
91503

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗