Open original FDA record ↗Recall record
- FDA identifier
- Z-1293-2024
- Company in FDA record
- Retractable Technologies, Inc.
- FDA classification date
- Mar 14, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The needle cannula of a retractable needle may detach from the needle holder in the patient.
Source facts
- Action
- On 2/5/24, recall notices were sent to customers and distributors who were asked to do the following: 1) Segregate, quarantine, and discontinue use and distribution of affected devices. 2) Inform all personnel who have received or are receiving affected devices that they should not use affected devices. 3) If affected product was further distributed, notify these additional customers immediately. When notifying customers include a copy of the recall notice, label images, and the response form. 4) Complete and return the response form to rtiservice@retractable.com Customers with additional questions can email the firm at regulatory@retractable.com
- Initiated
- 2024-02-05
- Root cause
- Process control
- Distribution
- US Nationwide distribution including in the states of NY, FL, IL, MA, UT, NJ, AZ, MO, WA, NC, CA, TX, NV, IN, PA, VA, IA, GA, TN, MI, OH, OK, CO, OR, SC, MN, CT.
- Recall number
- Z-1293-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206047
- FDA classified
- 2024-03-14
- Recall event number
- 93981