Reason stated in the source
Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
Source facts
- Action
- Medica issued letters dated 2/8/19 via email Bulletin B1901 & B1902 to distributors and end-users on record. Letter describes problem, action to be taken: Immediately discontinue use of the Wash1 wedges (REF 10680) and discard any remaining kits/wedges. Complete the response form. Discuss possible concerns about Lipase results achieved using these lots of Wash wedges with your laboratory manager and/or clinician in order to decide on appropriate measures. Report any concerns to Medica for evaluation and review. Questions contact Medica Technical Support at (877) 777-5895 or via email at: EasyRAtech@medicacorp.com.
- Initiated
- 2019-02-06
- Root cause
- Process design
- Terminated
- 2020-10-02
- Distribution
- Worldwide distribution - US Nationwide and countries of IRAQ, FRANCE, INDIA, GEORGIA, GEORGIA, FRANCE GUATEMALA, KENYA, IRAN, ESTONIA.
- Recall number
- Z-1294-2019
- Recall status
- Terminated
- CDRH record ID
- 171152
- FDA record created
- 2019-05-07
- Recall event number
- 82217