Z-1294-2022: Artis Pheno, Interventional Fluoroscopic X-Ray System, Model No. 10849000

Recall record Z-1294-2022 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1294-2022
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jun 27, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A video signal from one of the transceivers of the Large Display may be lost. In very rare cases, both transceivers of the Large Display could potentially fail which can lead to the loss of both video signals. This issue can sporadically affect the Large Display in the examination room as well as the displays in the control room.

Source facts

Action
Firm notified customers of the recall on June 10, 2022 via Urgent Medical Device Correction letters. The firm strongly recommends reinforcing appropriate emergency procedures until the corrective action has been performed. Please ensure that patient treatment can be continued in other ways if there is any possible danger to the safety of the patient. The firm will exchange the affected video transmission links. Their service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact the firm's service organization for an earlier appointment at 1-800-888-7436. The firm does not consider it necessary to re-examine any patients previously treated using the Artis pheno system.
Initiated
2022-06-10
Root cause
Process control
Distribution
Domestic distribution nationwide. Foreign distribution to Australia Austria Belgium Brazil Canada China Colombia Denmark Finland France Germany Hungary Indonesia Iran Italy Japan Kuwait Mexico Monaco Netherlands Norway Poland Singapore Slovakia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom United States Vietnam
Recall number
Z-1294-2022
Recall status
Open, Classified
CDRH record ID
194621
FDA classified
2022-06-27
Recall event number
90507

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗