Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists
Source facts
Action
Siemens issued Customer Safety Advisory Letter dated 3/22/19 advising of problem, health risk and action to take: recommends to avoid overwriting the organ programs, as this can lead to frequent operations. Siemens is distributing this letter and the addendum to the user manual to all potentially affected users Contact our service organization for any additional questions at 1-800-888-7436.
Initiated
2019-03-22
Root cause
Device Design
Terminated
2020-05-12
Distribution
US Nationwide Distribution in the states of NC, MA
Recall number
Z-1295-2019
Recall status
Terminated
CDRH record ID
172036
FDA record created
2019-05-07
Recall event number
82613