Z-1295-2021: VITROS Chemistry Products Calibrator Kit 4-(Calibrator values for TIBC only). Used to calibrate VITROS 2

Recall record Z-1295-2021 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1295-2021
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA record creation date
Mar 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential total iron-binding capacity (TIBC) results. The VITROS TIBC performance shift will start with the newest Generation (GEN) of VITROS Fe Slides (GEN 26), the potential shift in TIBC patient values of GEN 26 and above will be approximately 80 ug/dL lower across the measuring range than other available GENS of VITROS Fe Slides (GENs 22-25).

Source facts

Action
Ortho Clinical Diagnostics on 18 February 2021 issued Product Correction Notice(Ref. CL2021-054) states reason for correction, health risk and provided the following information: VITROS TIBC results will show a performance shift will start with the newest Generation (GEN) of VITROS Fe Slides (GEN 26) and above. Specifically, for VITROS TIBC GEN 26 and above, patient and quality control values may shift lower (approximately 80 g/dL across the measuring range) than other available GENS of VITROS Fe Slides (GENs 22-25). To maintain consistent VITROS TIBC GEN to GEN performance, Ortho will be updating the VITROS Calibrator Kit 4 TIBC values for VITROS Fe Slides GENs 22-25 to match VITROS Fe Slides GENs 26 and above. This will allow the TIBC performance for all VITROS Fe Slides GENs to match the reference method performance utilizing the new MgCO3 lot. The performance adjustment for VITROS Fe Slides GENs 22-25 is expected to be implemented in March 2021. At that time, VITROS TIBC GENs 22-25 will become uncalibrated. Ortho will notify customers in advance of this change. After noting that result of VITROS TIBC GEN 26 and above will shift approximately 80 g/dL from current performance, it is acceptable to continue using all VITROS Fe Slides GENs with the VITROS TIBC Kit and VITROS Calibrator Kit 4. " Complete the enclosed Confirmation of Receipt form no later than February 26, 2021 " Please forward this notification if the product was distributed outside of your facility If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Initiated
2021-02-18
Root cause
Process design
Terminated
2022-10-18
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
Recall number
Z-1295-2021
Recall status
Terminated
CDRH record ID
185946
FDA record created
2021-03-23
Recall event number
87390

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗