Skip to content

Public FDA record

Z-1295-2022: regard Item Number: 800774001, CV00979A - CV Minor Kit, Sterile.

Recall record Z-1295-2022 for ROi CPS LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1295-2022
Company in FDA record
ROi CPS LLC
FDA classification date
Jun 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A nonsterile component was packaged in a sterile convenience kit.

Source facts

Action
The recalling firm called the consignee on 5/20/2022 to discuss the recall, including quarantine of affected product, customer fulfilment/available inventory, end user facility notification, recall timelines, and establishing a plan for relabeling by the recalling firm personnel at the distribution center. The recalling firm issued an email to the consignee on 5/20/2022 with an attached recall notice that was previously discussed with the consignee. The email requested the consignee to quarantine the kits and immediately notify the end user facilities who have been shipped the affected kits. In the email, the recalling firm reported they would be sending someone to the direct account and to the end user facilities to complete the correction of the affected kits. The attached letter dated 5/20/22 informed the customer of the reason for recall, health risk assessment, and the actions to be taken by customers. The letter contained a picture of the non-sterile Ultrasound Gel that is to be discarded from the kits. A Recall Reply Form was enclosed to indicate the quantity of product in the customer's possession. The customer was also supposed to indicate that they have read and understand the notice and it has been forwarded to all customers who have been shipped the affected product.
Initiated
2022-05-20
Root cause
Process change control
Terminated
2023-05-02
Distribution
US Nationwide distribution in the state of FL.
Recall number
Z-1295-2022
Recall status
Terminated
CDRH record ID
194092
FDA classified
2022-06-28
Recall event number
90302

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗