Skip to content

Public FDA record

Z-1296-2021: Operating Table System

Recall record Z-1296-2021 for Getinge Group Logistics America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1296-2021
Company in FDA record
Getinge Group Logistics America, LLC
FDA record creation date
Mar 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.

Source facts

Action
In February, 2021, Maquet GmbH/Getinge, will issue an Urgent Medical Device Correction notice to customers via FedEx 2 Day Delivery for the Magnus 1180 Operating Table due to a potential for hydraulic leakage to occur at the inclination function which may result in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. If this were to occur, it may lead to the patient having to be transferred to another OR table causing a delay in the procedure.
Initiated
2021-02-11
Root cause
Device Design
Distribution
Domestic: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV. Foreign: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Luxembourg, Lybia, Macedonia, Malaysia, Mauritius, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom, Uzbekistan, Venezuela, Vietnam and Yemen.
Recall number
Z-1296-2021
Recall status
Open, Classified
CDRH record ID
185942
FDA record created
2021-03-23
Recall event number
87386

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗