Skip to content

Public FDA record

Z-1298-06: CONTAK RENEWAL TR (models H120, H125) and CONTAK RENEWAL TR2 (models H140, H145). CONTAK RENEWAL TR2 dev

Recall record Z-1298-06 for Boston Scientific CRM Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1298-06
Company in FDA record
Boston Scientific CRM Corp
FDA classification date
Sep 28, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Guidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion without warning in the affected devices.

Source facts

Action
A Press Release was issued on 06/26/06. An Urgent Medical Device Safety Information & Corrective Action letter, dated 06/23/06 was delivered to physicians describing preliminary information before Guidant has completed its own investigation and finalized patient care recommendations. Non-implanted inventory is being retrieved from hospital shelves and Guidant Represenatatives. Physicians are to perform an in-clinic follow-up exam as soon as possible with patients implanted with affected devices.
Initiated
2006-06-24
Root cause
Other
Terminated
2008-03-02
Distribution
Worldwide- USA and Antigua Barbuda, Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Micronesia, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Peru, Philippines, Polynesia, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Venezuela
Recall number
Z-1298-06
Recall status
Terminated
CDRH record ID
46735
FDA classified
2006-09-28
Recall event number
35754

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗