Continuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of continuous use. As a result, risk of premature failure of the Monoblock and loss of imaging X-rays.
Source facts
Action
Urgent Medical Device Correction letters dated 3/18/19 were sent to customers.
Initiated
2019-03-18
Root cause
Device Design
Terminated
2020-04-16
Distribution
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
Recall number
Z-1298-2019
Recall status
Terminated
CDRH record ID
172107
FDA record created
2019-05-08
Recall event number
82621