When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
Source facts
Action
On April 22, 2022, the firm notified affected customers of the recall via email.
Customers were asked to complete and return the attached acknowledgement form. Upon receipt, Haemonetics will contact the customer to arrange the remote installation of the updated software, version 4.4.0, which will resolve the issue.
Initiated
2022-04-19
Root cause
Software design
Terminated
2024-03-27
Distribution
US Nationwide.
Recall number
Z-1298-2022
Recall status
Terminated
CDRH record ID
194215
FDA record created
2022-06-28
Recall event number
90358