Skip to content

Public FDA record

Z-1298-2026: Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10

Recall record Z-1298-2026 for Wilson-Cook Medical Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1298-2026
Company in FDA record
Wilson-Cook Medical Inc.
FDA classification date
Feb 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Nasal feeding tube packaged without the nasal transfer tube component.

Source facts

Action
On January 5, 2026, Cook Endoscopy issued a Urgent Medical Device Recall Notification to affected consignees via E-Mail. Cook Endoscopy asked consignees to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s). If so, quarantine any affected products that remain unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a signed copy of the Acknowledgement and Receipt Form. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand and email the completed form to ausrecalls@cookmedical.com. 4. Please share this notice with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected products have been transferred.
Initiated
2025-12-15
Root cause
Employee error
Distribution
International distribution to the country of Australia.
Recall number
Z-1298-2026
Recall status
Open, Classified
CDRH record ID
218095
FDA classified
2026-02-05
Recall event number
98274

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.