Skip to content

Public FDA record

Z-1299-2022: RayStation 11B

Recall record Z-1299-2022 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1299-2022
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Jun 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.

Source facts

Action
The firm notified customers of the recall via email on June 10, 2022. The notice was titled "Field Safety Notice, Medical Device Correction #99834." Customers were given precautions and instructed on a workaround. The issue will be resolved in the next version of RayStation, scheduled for market release in June 2022, though this may be subject to market clearance.
Initiated
2022-06-10
Root cause
Under Investigation by firm
Terminated
2024-06-11
Distribution
US Nationwide distribution in the states of NJ and TX.
Recall number
Z-1299-2022
Recall status
Terminated
CDRH record ID
194605
FDA record created
2022-06-28
Recall event number
90505

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗