Z-1299-2026: FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Recall record Z-1299-2026 for FUJIFILM Healthcare Americas Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1299-2026
Company in FDA record
FUJIFILM Healthcare Americas Corporation
FDA classification date
Feb 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.

Source facts

Action
On January 9, 2026 URGENT: MEDICAL DEVICE RECALL letters were emailed to customers. Actions to be taken: Please pay attention to the following points until the work is completed. " For safety reasons, please turn off the breaker of the distribution board to which the power cord of the unit is connected after turning off the power of the unit according to the operation manual. " Do not forget to turn off the breaker of the distribution board until the replacement of the contactor is completed. 1) Please post this notification on or near the device 2) Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 3) Please contact Technical Assistance Center (TAC) at 1-888-FUJIMED (1-888-385-4633) to open a service ticket. 4) Please report any known adverse events to FDA via MedWatch.
Initiated
2026-01-09
Root cause
Process control
Distribution
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Recall number
Z-1299-2026
Recall status
Open, Classified
CDRH record ID
217665
FDA classified
2026-02-05
Recall event number
98124

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗