Z-1300-2026: Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

Recall record Z-1300-2026 for Maquet Cardiopulmonary Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1300-2026
Company in FDA record
Maquet Cardiopulmonary Gmbh
FDA classification date
Feb 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.

Source facts

Action
On or around January 9, 2026, "URGENT MEDICAL DEVICE RECALL" letters were sent to customers. Customers were instructed to take the following actions: - Please examine your inventory immediately to determine if you have any affected product in your inventory and utilize the recommended mitigations listed above until replacement sensors are available. - Complete and return the enclosed response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com. - Contact Customer Service via email at ACTSalesSupport.US@Getinge.com or 888-943-8872, ext 28849 / (888 9GETUSA), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. (Eastern Time Zone) to order a replacement sensor that will be shipped once available. Affected sensors will be returned to Getinge at time of replacement. - Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. If you have any questions, contact Customer Service via email at ACTSalesSupport.US@Getinge.com or 888-943-8872, ext 28849 / (888 9GETUSA), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. (Eastern Time Zone).
Initiated
2026-01-09
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of (South Korea), Latvia, Lithuania, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, and Vietnam
Recall number
Z-1300-2026
Recall status
Open, Classified
CDRH record ID
217747
FDA classified
2026-02-05
Recall event number
98159

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗