Open original FDA record ↗Recall record
- FDA identifier
- Z-1301-2016
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Mar 31, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio licenses "Cardio Pro" and "Cardio Retro".
Source facts
- Action
- A Safety Advisory Notice was sent to customers informing them of a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems. Siemens informed them that they will deliver an Update Instruction for their Service Engineers to modify the installed Firmware. This will correct the rare synchronization errors of Scanner Firmware and Scanner Software to start the application of the x-ray in compliance with the optimum triggering point for best image quality.
- Initiated
- 2016-02-15
- Root cause
- Device Design
- Terminated
- 2016-12-07
- Distribution
- US (nationwide) including Guam and the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, VA, and WA.
- Recall number
- Z-1301-2016
- Recall status
- Terminated
- CDRH record ID
- 144268
- FDA record created
- 2016-03-31
- Recall event number
- 73477