A software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving corrected/modified values. Outdated information could be sent to the referring physician when creating the formal report in the reporting system.
Source facts
Action
Customer Advisory Notice SY065/17/P was sent to direct accounts. Firm is continuing to work on software solution. A workaround is available and is described in the Customer Advisory Notice.
Initiated
2018-01-08
Root cause
Software design
Terminated
2020-04-22
Distribution
Device is software only. No products are distributed to wholesale dealers, distributers or retailers
Recall number
Z-1303-2018
Recall status
Terminated
CDRH record ID
162599
FDA classified
2018-03-20
Recall event number
79532