Z-1303-2023: Express Mini 500 Dry Seal Chest Drain, Part Number 16400

Recall record Z-1303-2023 for Atrium Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1303-2023
Company in FDA record
Atrium Medical Corporation
FDA classification date
Mar 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Instructions for Use (IFU) for the Pneumostat Chest Drain Valve and the Express Mini 500 Dry Seal Chest Drains do not provide sufficient precaution against draining the device for continued use. Additionally, the IFUs do not provide sufficient precaution or warning against the use of the device outside the healthcare setting.

Source facts

Action
Urgent Medical Device Correction letters dated 3/1/23 were sent to customers. Updated Instructions New Warning and Precautions: The facility can continue use of the product with the IFU that was provided along with the consideration of the following: " NEW and Applicable Existing Warnings and Precautions for Express Mini 500: o NEW Warning: The Express Mini 500 is restricted for use by trained healthcare providers familiar with cardiothoracic surgical procedures and techniques, including the use of chest drains. o NEW Precaution: The Express Mini 500 is restricted for use in a healthcare facility. The Express Mini 500 should not be used for outpatient drainage. o NEW Precaution: Use of the Luer port is intended only for sampling patient drainage. Do not use the Luer port or any other means to empty fluid from the collection chamber. o Existing Precaution: Replace chest drain if damaged or when collection volume meets or exceeds maximum capacity. Actions to be Taken by the Customer: " Please ensure that all Atrium Express Mini 500 Dry Seal Chest Drain users at your facility are aware of this Safety Notice and post a copy of the Notice on Page 3 in all inventory locations within your facility where the devices are stored. Your facility can continue use of the device. No devices need to be returned. " Please forward this information to all current and potential Atrium Express Mini 500 Dry Seal Chest Drain users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. " Please complete and sign the attached URGENT: MEDICAL DEVICE CORRECTION - RESPONSE FORM on Page 4 to acknowledge that you have received this notification. Return the completed form to Atrium/Getinge by emailing a scanned copy of the completed form to Mini500-Pneumostat-2023.Atrium@getinge.com or or by faxing the form to (866)-308-8040. Type of Action Taken by Getinge: Atrium
Initiated
2023-03-01
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Austria, Australia, Belgium, Brazil, Canada, Switzerland, Colombia, Czech Republic, Germany, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Indonesia, Ireland, Israel, Iceland, Italy, Jordan, South Korea, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Peru, Philippines, Poland, Portugal, Sweden, Singapore, Slovenia, Thailand.
Recall number
Z-1303-2023
Recall status
Open, Classified
CDRH record ID
198707
FDA classified
2023-03-30
Recall event number
91718

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗