Public FDA record

Z-1305-2025: BD Care Coordination Engine (CCE) Infusion Adapter, Safety Management Software

Recall record Z-1305-2025 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1305-2025
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Mar 13, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).

Source facts

Action
On February 18, 2025, Becton, Dickinson issued a " Urgent Medical Device Product Correction" Notification. to affected consignees via Fed-Ex and E-Mail. 1. Users should review all infusion parameters pre-populated and confirm they are correct before pressing START as stated in the BD Alaris User Manual. 2. Disseminate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 3. BD Project Managers will engage with you to address any questions regarding this notice, and specific next steps for your facility. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form, regardless of whether you have any of the impacted material. This will allow BD to confirm receipt of this notification. 5. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program.
Initiated
2025-02-18
Root cause
Under Investigation by firm
Distribution
US: AZ, CA, DC, FL, GA, IL, IN, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI OUS: None
Recall number
Z-1305-2025
Recall status
Open, Classified
CDRH record ID
212552
FDA classified
2025-03-13
Recall event number
96323

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗