Failure of Data Transmission: No imaging data could be transmitted by the device, because a key value needed for data transmission was incorrectly entered in the affected units by AGFA Healthcare.
Source facts
Action
On 12/11/07, AGFA Healthcare sent out an e-mail to all affected consignees informing them of the situation. The firm also arranged to install the corrected License Key on all affected units of the product.
Initiated
2007-12-11
Root cause
Release of Material/Component prior to receiving test results
Terminated
2010-07-21
Distribution
Nationwide Distribution - USA, including states of CA, GA, IA and SD.
Recall number
Z-1306-2008
Recall status
Terminated
CDRH record ID
68301
FDA classified
2008-07-01
Recall event number
46812