Reason stated in the source

An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.

Source facts

Action
The firm disseminated an Urgent Medical Device Correction Notice beginning on 05/31/2022 by letter. The letter described the problem and requested the following actions be taken by the customer: "Identify if your Nurse Call System utilizes one of the impacted wireless integrations identified. If so, please share the information with potential users in your organization. Hillrom recommends reviewing the wireless system settings with your wireless provider to determine if there are configurable options for call acceptance only. Complete the enclosed response form and return to hillromCARY021@sedgwick.com within two weeks of receipt of this notice, to confirm you have received this communication and understand the potential issue." Persons with questions were directed as follows: "If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support, available twenty-four hours a day, seven days a week, at 1-800-445-3720, option 3, or hrccwssupport@hillrom.com"
Initiated
2022-05-31
Root cause
Software Design Change
Distribution
nationwide and Canada
Recall number
Z-1306-2022
Recall status
Open, Classified
CDRH record ID
194130
FDA record created
2022-06-30
Recall event number
90322

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗