Open original FDA record ↗Recall record
- FDA identifier
- Z-1306-2022
- Company in FDA record
- Baxter Healthcare Corporation
- FDA record creation date
- Jun 30, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.
Source facts
- Action
- The firm disseminated an Urgent Medical Device Correction Notice beginning on 05/31/2022 by letter. The letter described the problem and requested the following actions be taken by the customer: "Identify if your Nurse Call System utilizes one of the impacted wireless integrations identified. If so, please share the information with potential users in your organization. Hillrom recommends reviewing the wireless system settings with your wireless provider to determine if there are configurable options for call acceptance only. Complete the enclosed response form and return to hillromCARY021@sedgwick.com within two weeks of receipt of this notice, to confirm you have received this communication and understand the potential issue." Persons with questions were directed as follows: "If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support, available twenty-four hours a day, seven days a week, at 1-800-445-3720, option 3, or hrccwssupport@hillrom.com"
- Initiated
- 2022-05-31
- Root cause
- Software Design Change
- Distribution
- nationwide and Canada
- Recall number
- Z-1306-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194130
- FDA record created
- 2022-06-30
- Recall event number
- 90322