Z-1307-2026: Product Name: Lumipulse -Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applic

Recall record Z-1307-2026 for Fujirebio Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1307-2026
Company in FDA record
Fujirebio Diagnostics, Inc.
FDA classification date
Feb 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Source facts

Action
On or about December 11, 2025 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Immediately discontinue using the associated Lots listed above for the Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio. - Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. - Notify your ordering physician or health care provider as appropriate. - The test results must be interpreted in conjunction with other diagnostic tools and clinical information. - Please retain this letter with your laboratory records and forward this leer to those who may have received this product.
Initiated
2025-12-11
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Recall number
Z-1307-2026
Recall status
Open, Classified
CDRH record ID
217946
FDA classified
2026-02-05
Recall event number
98232

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗