Philips Medical Systems Sales & Service Region No. America
FDA classification date
Aug 4, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
When reviewing images in the module before sending them to PACS or printer the order can be erratic and out of sequence and may cause confusion with the interpretation of the examination.
Source facts
Action
The firm mailed letters via certified mail to their affected customers beginning on June 8, 2004. The letters instruct their customers to modify the default display on the ViewForum to exam/series date and time until a service representative can install the upgraded software R3.2.
Initiated
2004-06-08
Root cause
Other
Terminated
2006-02-09
Distribution
Philips Medical Systems North American Company distributes product to medical facilities nationwide.
Recall number
Z-1308-04
Recall status
Terminated