The gradient output supervision was permanently turned off on the MAGNETOM system, meaning that gradient outputs could exceed IEC60601-2-33 limits and peripheral nerve stimulation could occur.
Source facts
Action
A Customer Safety Advisory Notice, dated March 6, 2015, was sent to end users to inform them of the potential issue and what measures were being taken to mitigate possible risks.
Initiated
2015-03-09
Root cause
Nonconforming Material/Component
Terminated
2016-08-17
Distribution
Nationwide Distribution.
Recall number
Z-1308-2015
Recall status
Terminated
CDRH record ID
134534
FDA classified
2015-03-20
Recall event number
70712
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.