Z-1308-2016: VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator
MHI-TM2000 Linear
Recall record Z-1308-2016 for MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK. View source facts, labeled dates, and the original FDA record.
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
FDA classification date
Mar 31, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is incorrectly set and if nobody is aware of it, an erroneous treatment would be given to the patient.
Source facts
Action
A Field Safety Notice dated March 2, 2016, was sent to all affected customers via e-mail on March 7, 2016. The Field Safety Notice informed users of the problem and provided a safe workaround until corrected software is available. The correction will be free of charge and a Customer Service Representative will contact all affected customers to schedule its installation. Customers with questions were advised to contact their Customer Service Representative.
Initiated
2016-03-07
Root cause
Software design
Terminated
2017-04-20
Distribution
Nationwide Distribution including New York, Texas, Florida, and Ohio
Recall number
Z-1308-2016
Recall status
Terminated
CDRH record ID
144310
FDA classified
2016-03-31
Recall event number
73436