Open original FDA record ↗Recall record
- FDA identifier
- Z-1308-2021
- Company in FDA record
- Biomerieux Inc
- FDA record creation date
- Mar 25, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).
Source facts
- Action
- The customer letter instructs the customer on how to verify the external communication configuration setting to see if they are impacted by this correction. If impacted, the customer will need to contact their local bioMerieux representative for instructions on how to verify expertised results until a correction is implemented. For a correction, customers can either convert to MYLA or BCI (where this issue will not occur) or continue to verify the expertised results until a software patch is deployed for 9.0x customers.
- Initiated
- 2021-01-28
- Root cause
- Software design
- Terminated
- 2023-01-12
- Distribution
- Worldwide distribution - No US consignees affected and the countries of Andorra, Austria, Bangladesh, Bermuda, Brazil, Brunei Darussalam, El Salvador, French Guiana, French Polynesia, Gibraltar, Greece, Guadeloupe, Guam, Holy See (Vatican City State), Hong Kong, Hungary, India, Indonesia, Ireland, Lao People's Democratic Republic, Luxembourg, Macao, Malaysia, Martinique, Mexico, Myanmar, New Caledonia, New Zealand, Nigeria, Philippines, Reunion, San Marino, Slovakia, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Uruguay.
- Recall number
- Z-1308-2021
- Recall status
- Terminated
- CDRH record ID
- 185827
- FDA record created
- 2021-03-25
- Recall event number
- 87249