Z-1308-2022: VOLISTA StandOP Surgical Light, Product Code/Part Numbers ARD568811901 VLT600SF AIM STP ARD568811911 VLT6

Recall record Z-1308-2022 for Getinge Usa Sales Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1308-2022
Company in FDA record
Getinge Usa Sales Inc
FDA classification date
Jun 30, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Paint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical operation into the patient body or in the operating field.

Source facts

Action
On May 6, 2022, the firm notified affected customers through an Urgent Medical Device Correction letter. Customers were instructed to continue to perform daily inspections per the IFU, which include checking for any chipped or missing paint. If you observe paint chipping or detachment, the firm recommends that you stop using the device and contact your Getinge service representative to schedule an appointment to replace the part free of charge. The firm is requesting that customers contact their Getinge service representative to have the part replaced if chipped or missing paint is observed through April 2025. Getinge service representatives will inspect the affected component during periodic maintenance and will replace the component if the issue is observed through April 2025.
Initiated
2022-05-06
Root cause
Process change control
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, GA, IA, ID, KS, KY, LA, MA, MD, ME, MO, NE, NY, OH, OR, TN, TX, UT, VA, WA.
Recall number
Z-1308-2022
Recall status
Open, Classified
CDRH record ID
194099
FDA classified
2022-06-30
Recall event number
90305

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗