When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As
a result there is the potential to deliver dose to the initial isocenter rather than the new location.
Source facts
Action
ViewRay sent an Urgent Device Correction letter dated March 7, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Customers with questions were instructed to contact ViewRay Customer Support at support@viewray.com or call the ViewRay Customer Support center at 855-286-8875.
Initiated
2016-03-07
Root cause
Software design
Terminated
2017-02-22
Distribution
Worldwide Distribution - US including CA, FL, MO & WI., and International distribution to the Republic of Korea and The Netherlands.
Recall number
Z-1309-2016
Recall status
Terminated
CDRH record ID
144447
FDA classified
2016-03-31
Recall event number
73550