- Recall classification
- Class II
Reason stated in the source
Under certain circumstances, the mounting screws may loosen or pull out from the attachment point on the Ergon 3 boom when the GCX Channel Mounting Accessory was used with the VST clamp as a monitor mount on the boom.
Source facts
- Action
- On June 2, 2022, the firm sent out Urgent Medical Device Correction letters to affected customers. Customers were instructed to identify their GCX Channel Mounting Accessories used with Ergon 3 booms. Customers should inspect their mounting accessories. If the user sees that the channel is tilted on the VST clamp, this could indicate that the screws or tension knob have loosened. Remove the device from service if you find evidence that it may fail. Skytron authorized representatives will contact customers to schedule a service visit. Please contact quality at quality@skytron.com with any questions.
- Initiated
- 2022-06-02
- Root cause
- Component design/selection
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada and Japan.